Reported Apr 18, 2018 · Terminated

VITROS Immunodiagnostic Products TSH Reagent Pack

Clinical Diagnostic Systems

Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Clinical Diagnostic Systems · 601 Lee Rd Ny · Rochester, NY · 14652-0001

Is this the product you have?

  • VITROS Immunodiagnostic Products TSH Reagent Pack
Lot codes and dates
  • Lot 5470
  • expiry date24-May-2018 Lot 5430
  • expiry date01-Mar-2018 (not distributed in the US.) Lot 5178 was also affected by this issue, expired on 31-Mar-2017
Amount recalled
Lot #5470 (US=8570/OUS=537); Lot #5430 (US=0/OUS=9135); Lot #5178 (US=5178/OUS=4396)

Why it was recalled

VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside of the calibration quality parameter measuring the gradient of the Customer Calibration Curve between Calibrator Level 2 and Calibrator Level 3.

This product is unlikely to cause harm, but it breaks a safety or labelling rule.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Clinical Diagnostic Systems recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 8, 2020 · Data as of Sep 19, 6:12 PM UTC