Reported Feb 26, 2020 · Ongoing
CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic,…
Philips Medical Systems GmbH, DMC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Medical Systems Gmbh, DMC · Development And Manufacturing Ctr. · Hamburg · N/A
Is this the product you have?
- CombiDiagnost GCF, Software v. 1.0.5
- Product Code 709030 - Product UsageXray system, diagnostic, fluoroscopic, general purpose
- Lot codes and dates
- (domestic only)
SN18061848SN17071689SN18051832SN190251SN17111759SN17111752SN1801179310000020SN17121783SN17121785SN18021816SN18101890SN19031950SN17071688SN18101903SN17071683SN17101722SN17101729SN17111758SN17111763SN17031622SN17071692SN17041642SN18101886SN18011794SN19021934SN19021943SN17111770SN18011802SN19021942SN17121779SN17031618SN17101733SN17021613SN17021613
- (domestic only)
- Amount recalled
- 116
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Medical Systems GmbH, DMC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.