Reported May 22, 2019 · Terminated
Terumo Advanced Perfusion System 1 Electronic Patient Gas System, Catalog # 801188, UDI 00886799000588 The electronic…
Terumo Cardiovascular Systems Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Terumo Cardiovascular Systems Corporation · 6200 Jackson Rd · Ann Arbor, MI · 48103-9586
Is this the product you have?
- Terumo Advanced Perfusion System 1 Electronic Patient Gas System, Catalog # 801188, UDI 00886799000588 The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit.
- Lot codes and dates
- All
- Amount recalled
- 1258
Where it was sold
- Nationwide
Why it was recalled
The gas flow rate output of the EPGS may be inaccurate due to a defect in the internal flowmeter.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Terumo Cardiovascular Systems Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.