Reported Apr 13, 2016 · Terminated

SIGNA Pioneer MR Systems Product Usage: The SIGNA Pioneer is a whole body magnetic resonance scanner designed to…

GE Medical Systems, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Medical Systems, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

  • SIGNA Pioneer MR Systems Product UsageThe SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Pioneer reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis
Lot codes and dates
  • Mfg Lot or Serial # System ID 001612TA1 N/A
    • 001718TA6
    • 336397P3T
    • 002331TA7
    • 083132425984
    • 001719TA4
    • 250598MR1
    • 001424TA1
    • M2850013
    • 001537TA0
    • TP0008MR07
    • 001662TA6
    • 157022MR01
    • 001184TA1
    • ZC0002
    • 001652TA7
    N/A 002227TA7 ZC0003 002278TA0 ZC0004 002330TA9 ZC0006 002529TA6 N/A 001485TA2 34026MRS05 002279TA8 34225MRS02 002395TA2 34185MRS02
Amount recalled
16 (2 US; 14 OUS)

Where it was sold

Why it was recalled

Ferrous material in the SIGNA Pioneer Table Control Box (TCB) could cause unexpected attraction to the MR system's high strength magnetic field during servicing activities. This could create a potential risk of serious injury to service personnel who attempt to inappropriately remove the TCB from the magnetic field. There have been no events reported as a result of this issue. Clinical scanni

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Medical Systems, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 16, 2016 · Data as of Sep 19, 6:05 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.