Reported Mar 4, 2020 · Terminated
THP Hip Plating System, Item Nos. 14-451070 14-451075 14-451080 14-451085 14-451090 14-451095 14-451100 14-451105…
Zimmer Biomet, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Zimmer Biomet, Inc. · 56 E Bell Dr · Warsaw, IN · 46582-6989
Is this the product you have?
- THP Hip Plating System, Item Nos.. The product is a metallic bone fixation system.
14-45107014-45107514-45108014-45108514-45109014-45109514-45110014-45110514-45111014-45111514-45112014-45112514-451130
- Lot codes and dates
- Lots
148 of 148 codes
397440875130018810018850018860018880018900030090030100030140030150030180049860049870049900096520108200108290129340129420129440129460129500129520129550129570129580129590129600129620142280142290142310348480397350397370397400397430407250407260407270434970434980527970527990552890553000606240606260606300613290613300613310613320613330613340613350613360613380613390
- Lots
- Amount recalled
- 1662
Where it was sold
Why it was recalled
Potential increase in fracture rate was detected during internal testing, in addition to potential discoloration inside the inner pouch of the packaging.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Zimmer Biomet, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.