Reported Mar 27, 2024 · Ongoing

MINI STICK MAX 5F X 15 CM STIFF .018 NI/TU NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into…

Angiodynamics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Angiodynamics, Inc. · 603 Queensbury Ave · Queensbury, NY · 12804-7619

Is this the product you have?

  • MINI STICK MAX 5F X 15 CM STIFF .018 NI/TU NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system
  • Catalog Number45-776
Lot codes and dates
  • UDI15051684023191
  • UPNH965457761
  • Lot Number5787840 5803520 5805069
Amount recalled
40 units

This recall covers 15 products

The agency filed one notice per product. You are reading one of them; the other 14 are below.

2 more are in the archive.

Where it was sold

  • Nationwide
  • CA

Why it was recalled

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Angiodynamics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:44 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.