Reported Mar 4, 2020 · Terminated
Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft.
Cook Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Cook Inc. · 750 N Daniels Way · Bloomington, IN · 47404-9120
Is this the product you have?
- Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft.
- Lot codes and dates
- All lots. OUS distribution only; Product is not cleared or distributed in the USA.
- Amount recalled
- 68,241
Why it was recalled
The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these factors. In addition, the firm has updated the Planning and Sizing Worksheet to include associated information.