Reported Apr 18, 2018 · Terminated
Normed Charcot Osteotome Chisel (Various sizes)
Zimmer GmbH
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Zimmer GmbH · Sulzerallee 8 · Winterthur · N/A
Is this the product you have?
- Normed Charcot Osteotome Chisel (Various sizes)
- Lot codes and dates
- Models26.08.506 OSTEO 8MM AND W/HDL 20CM/8IN 26.08.507 OSTEO 10MM AND W/HDL 20CM/8IN 26.08.508 OSTEO 12MM AND W/HDL 20CM/8IN 26.08.509 OSTEO 13MM CRVD R 10 23CM/9IN 26.08.619 OSTEO 19MM CRVD R 10 23CM/9IN (No US distribution) 26.08.712 OSTEO 12MM CRVD R 6 23CM/9IN (No US distribution) 26.08.715 OSTEO 15MM CRVD R 6 23CM/9IN (No US distribution) 26.08.718 CHARC OSTE 18MM CRD R6 23CM/9 26.08.812 OSTEO 12MM CRVD R 8 23CM/9IN (No US distribution) 26.08.815 OSTEO 15MM CRVD R 8 23CM/9IN (No US distribution) 26.08.818 CHARC OSTE 18MM CRD R8 23CM/9 (No US distribution) 26.08.900 ARTHROPIC ARTH 18CM 45TUFF HDL 26.08.910 RUTREK CHIS 20CM/8 DBL CUT BLD
Where it was sold
Why it was recalled
Product may contain a gap between the handle and metal chisel which can lead to failure to properly disinfect or sterilize between use.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Zimmer GmbH recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.