Reported Apr 20, 2016 · Terminated

Neurotherm NT2000iX RF Generator Model Number: RFG-NT -2000iX

Neurotherm, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Neurotherm, Inc. · 600 Research Dr Ste 1 · Wilmington, MA · 01887-4438

Is this the product you have?

  • Neurotherm
  • NT2000iX RF Generator Model NumberRFG-NT -2000iX
Lot codes and dates
  • Serial Numbers
    • 2593-13
    • 2717-13
    • 2616-13
    • 2704-13
    • 2735-13
    • 2360-12
    • 2732-13
    • 2727-13
    • 2284-11
    • 2726-13
    • 2728-13
    • 2426-12
    • 2700-13
    • 2508-12
    • 2643-13
    • 2733-13
    • 2730-13
    • 2724-13
    • 2453-12
    • 2631-13
    • 2171-11
    • 2059-10
    • 2729-13
    • 2294-12
    • 2725-13
Amount recalled
25

Where it was sold

Why it was recalled

NT2000iX software shipped with the international setting turned on. The international settings include access to Corodotomy, Bi-Polar, and No Temperature modes were not cleared in the United States for use.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Neurotherm, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 14, 2016 · Data as of Sep 19, 6:04 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.