Reported May 29, 2019 · Terminated

CHOL+HDL+GLU test strips, REF 2412

Polymer Technology Systems, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Polymer Technology Systems, Inc. · 7736 Zionsville Rd · Indianapolis, IN · 46268-2175

Is this the product you have?

  • CHOL+HDL+GLU test strips, REF 2412
Lot codes and dates
  • I808 I809 I811 I812 I814
Amount recalled
67643 total

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Polymer Technology Systems, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 24, 2020 · Data as of Sep 19, 6:53 PM UTC