Reported Apr 20, 2016 · Terminated

The Concerto Basic Shower Trolley is intended for assisted hygiene care, especially showering and bathing of residents…

Arjo, Inc. dba ArjoHuntleigh

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Arjo, Inc. dba ArjoHuntleigh · 2349 W Lake St · Addison, IL · 60101-6183

Is this the product you have?

  • The Concerto Basic Shower Trolley is intended for assisted hygiene care, especially showering and bathing of residents in care environments such as senior/assisted living, group home, special care, nursing homes, hospitals and home care. The Concerto Basic Shower Trolley is intended for indoor use.
Lot codes and dates
  • All Concerto & Basic Shower Trolleys manufactured by ArjoHuntleigh Polska Sp. z o.o, March 7, 2014 through April 20, 2015 with serial numbers from P0227559 - P0322393
Amount recalled
1,596 units

Where it was sold

Why it was recalled

Reports of the bolts connecting the stretcher to the hydraulic piston and the stretcher tilting mechanism becoming loose.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Arjo, Inc. dba ArjoHuntleigh recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 25, 2018 · Data as of Sep 19, 6:03 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.