Reported Mar 11, 2020 · Terminated
AIA-360 Automated Immunoassay Analyzer, Product Code 019945
Tosoh Bioscience Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Tosoh Bioscience Inc · 3600 Gantz Rd · Grove City, OH · 43123-1895
Is this the product you have?
AIA-360 Automated Immunoassay AnalyzerProduct Code 019945
- Lot codes and dates
284691022862920528469202286293052846930228629405284694022862950528469502286296052846960228629705284697022870990728469802287190072846990228719107284790022871920728519303287193072851940328719407285195032871950728519603287196072851970328719707285198032871980728519903287199072852900328729007285291032872910828529203287292082852990328729308285390032872940828539103287295082853920328729608285393032875990828539403287690082853950328769108285396032876920828539703287693082853980328769408
- Amount recalled
- 12
Where it was sold
Why it was recalled
A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Tosoh Bioscience Inc. recalls