Reported Jun 5, 2013 · Terminated

AU2700 Clinical Chemistry Analyzer AU5400 Clinical Chemistry Analyzer Perform automated analysis of serum, urine, and…

Beckman Coulter Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232

Is this the product you have?

  • AU2700 Clinical Chemistry Analyzer AU5400 Clinical Chemistry Analyzer Perform automated analysis of serum, urine, and other body fluids, including whole blood.
Lot codes and dates
  • All Serial Numbers Model Numbers
    • AU2700
    • AU27000ISE
    • AU5421
    • AU5421ISE
    • AU5431
  • Part Numbers
    • E0750053
    • N3663400
    • N3663700
    • N3664000
    • N3664200
    • N3664400
    • N3664600
    • N3664800
    • N3878500
    • 1465600
    • 64270G
    • E61DAU2700
    • N3663500
    • N3663600
    • N3663800
    • N3663900
    • N3664100
    • N3664300
    • N3664500
    • N3664700
    • N3664900
    • N3878600
    • O9102700
    • N3665000
    • N3665100
    • N3665200
    • N3665300
    • N3665400
    • N3878300
    • 64542
    • A94504
    • N1683100
    • AU5431-U1
    • N3665500
    • N3665600
    • N3665700
    • N3665800
    • N3665900
    • N3878400
Amount recalled
108 units

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400 Clinical Chemistry Analyzers becuase it was determined that the tubing/joint connections associated with these MODs may leak concentrated detergent. Beckman Coulter has received complaints that the joint connecting the tubing appears to disintegrate over time, causing a leak.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beckman Coulter Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 4, 2014 · Data as of Sep 19, 4:53 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.