Reported Apr 22, 2015 · Terminated

Magellan Robotic System; Model No(s). 11132 and 11139. A steerable catheter control system used at clinical sites.

Hansen Medical Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Hansen Medical Inc · 800 E Middlefield Rd · Mountain View, CA · 94043-4030

Is this the product you have?

  • Magellan Robotic System; Model No(s). 11132 and 11139. A steerable catheter control system used at clinical sites.
Lot codes and dates
  • All serial numbers.
Amount recalled
19 total installed systems

Why it was recalled

Based on investigation of two complaint incidents, all Hansen Medical Magellan Robotic System (Model No. 11132 and 11139) may experience an unexpected event during a retraction operation of catheter. Retraction of catheter may not stop when equipment button is released.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Hansen Medical Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 23, 2015 · Data as of Sep 19, 6:51 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.