Reported Jun 26, 2013 · Terminated
Philips BuckyDiagnost Radiographic. Catalog Numbers: 704031, 704032, 704035, 704060, 704062 This system is used for…
Philips Healthcare Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Healthcare Inc. · 3000 Minuteman Road · Andover, MA · 01810
Is this the product you have?
- Philips BuckyDiagnost Radiographic
- Catalog Numbers704031, 704032, 704035, 704060, 704062 This system is used for making x-ray exposures for diagnostics
- Lot codes and dates
- Serial Numbers
115 of 115 codes
334584345869349805352617375957/SN07000650352645355617356954363678362217362252361021364214362429361526363545363802364816366510366090367504370290370643373506372845376998383090384688/SN07000603386397/SN07000659406161/SN08000556386265/SN08000415389969/SN08000090381103/SN08000131444254/SN09000380406491/SN08000315389383/SN08000100447038/SN10000169389970/SN08000092406017/SN08000307389972/SN08000080389656/SN08000052386679/SN08000051390255/SN08000050394368/SN08000196396406/SN08000193399529/SN08000297401268/SN08000268402249/SN08000323399571/SN08000559384510/SN07000618402246/SN08000261403330/SN08000296403648/SN08000292407890/SN08000377405887/SN08000381413307/SN08000521413205/SN08000562411311/SN08000528419318/SN09000042428504/SN09000200
- Serial Numbers
- Amount recalled
- 116 units
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
Why it was recalled
An inspection of IATD (Installation Acceptance Test Document)IA SID (Source Image Distance) records shows table height measurements out of specification with the consequence of potential system mis-alignment
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Healthcare Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.