Reported Apr 16, 2014 · Terminated

Reusable latex breathing bags with multiple part numbers and sold as part of test lung part number and reusable…

Draeger Medical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Draeger Medical, Inc. · 3135 Quarry Rd · Telford, PA · 18969-1042

Is this the product you have?

  • Reusable latex breathing bags with multiple part numbers and sold as part of test lung part number and reusable anaesthesia sets
  • Product UsageThe breathing bag is a component of the anesthesia set (breathing circuit) that is intended to is the conduit for administering medical gases to a patient during ventilation. It provides both an inspiratory and expiratory route. 2. The test lung is an apparatus that contains a breathing bag and and connector. It is connected to the output of a ventilator and is used as part of the device/system check out
Lot codes and dates
  • Reusable latex breathing bags with part numbers 2165686, 2165953, 2165694, as part of test lung part number 8403201, and reusable anaesthesia sets part numbers M33681 and M27542.
Amount recalled
433

Where it was sold

Why it was recalled

Reusable latex breathing bags with various part numbers and sold as part of test lung and reusable anaesthesia sets were recalled when Draeger became aware of an incident where a user suffered an acute allergic reaction after using a Dr¿ger latex test lung. The user did not suffer permanent injury. The breathing bag that belongs to the test lung and other latex breathing bags are labelled NR. A

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Draeger Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 8, 2014 · Data as of Sep 19, 6:47 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.