Reported Mar 18, 2020 · Terminated

Monoject 3mL Syringe with Hypodermic Safety Needle, 22G x 1-1/2 . Item Code 11832215 - Product Usage: used for general…

Cardinal Health 200, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Cardinal Health 200, LLC · 15 Hampshire St Bldg 5 · Mansfield, MA · 02048-1113

Is this the product you have?

  • Monoject 3mL Syringe with Hypodermic Safety Needle, 22G x 1-1/2
  • Item Code 11832215 - Product Usageused for general purpose injection and aspiration of fluids from vials, ampules and parts of the body below the surface of the skin
Lot codes and dates
    • Lot No. 15072015
    • Item Code 11832215
    • Expires 06/20/2020
Amount recalled
8400 each

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

Why it was recalled

Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo sterilization process and is non-sterile. The usage of a needle that is not sterile could result in infection. Firm is not aware of any reports of patient harm.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Cardinal Health 200, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 4, 2020 · Data as of Sep 19, 5:33 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.