Reported Apr 23, 2014 · Terminated
Alligator Retrieval Device (ARD). Used in the peripheral and neurovasculature for foreign body removal.
Micro Therapeutics Inc., dba Ev3 Neurovascular
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Micro Therapeutics Inc, Dba Ev3 Neurovascular · 9775 Toledo Way · Irvine, CA · 92618-1811
Is this the product you have?
- Alligator Retrieval Device (ARD). Used in the peripheral and neurovasculature for foreign body removal.
- Lot codes and dates
- ARD Model/ Lot FA8881040 / 9769118 FA8881030 / 9816048 FA8881040 / 9754682 FA8881050 / 9772283 FA8881040 / 9752800 FA8881030 / 9751662 FA8881030 / 9773525 FA8881020 / 9832735 FA8881040 / 9757667 FA8881020 / 9774620 FA8881030 / 9834039 FA8881040 / 9758530 FA8881020 / 9774620 FA8881030 / 9834039 FA8881030 / 9758532 FA8881030 / 9783730 FA8881020 / 9836553 FA8881050 / 9760946 FA8881040 / 9786013 FA8881030 / 9834039 FA8881030 / 9761690 FA8881040 / 9787928 FA8881050 / 9848544 FA8881020 / 9761692 FA8881030 / 9791385 FA8881030 / 9848543 FA8881030 / 9758532 FA8881040 / 9793305 FA8881030 / 9853152 FA8881040 / 9764206 FA8881030 / 9798648 FA8881030 / 9854769 FA8881040 / 9764206 FA8881030 / 9798648 FA8881030 / 9858153 FA8881040 / 9764212 FA8881050 / 9801041 FA8881030 / 9865759 FA8881030 / 9766949 FA8881020 / 9811215 FA8881050 / 9868025 FA8881030 / 9769117 FA8881030 / 9813908
- Amount recalled
- 567 units total (323 units in US)
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
- PR
Why it was recalled
Stop using this product now
Potential for the PTFE (polytetrafluoroethylene) coating to delaminate and detach from the delivery wire.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Micro Therapeutics Inc., dba Ev3 Neurovascular recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.