Reported Mar 4, 2026 · Ongoing
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX,…
Agiliti Health - Ellis
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Agiliti Health - Ellis · 204 W 2nd St · Ellis, KS · 67637-2026
Is this the product you have?
- Adapt Pump
- Model/Catalog Number61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
- Lot codes and dates
- UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025
- Amount recalled
- 4286
Where it was sold
- Nationwide
Why it was recalled
Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Agiliti Health - Ellis recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.