Reported Mar 18, 2020 · Terminated

QIAGEN Rotor Gene Q Software version 2.3.4. Real-time PCR analyzer designed for rapid thermalcycling and real-time…

Qiagen Sciences LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Qiagen Sciences LLC · 19300 Germantown Rd · Germantown, MD · 20874-1415

Is this the product you have?

  • QIAGEN Rotor Gene Q Software version 2.3.4. Real-time PCR analyzer designed for rapid thermalcycling and real-time detection of PCR assays.
Lot codes and dates

6 of 6 products

Products covered by this recall
#Product as published
1refurbished 9001690 Rotor-Gene Q 2Plex HRM no PC 9001690R Rotor-Gene Q 2Plex HRM no PC
2refurbished 9001700 Rotor-Gene Q 5Plex no PC 9001700R Rotor-Gene Q 5Plex no PC
3refurbished 9001710 Rotor-Gene Q 5Plex HRM no PC 9001710R Rotor-Gene Q 5Plex HRM no PC
4refurbished 9001720 Rotor-Gene Q 6Plex no PC 9001720R Rotor-Gene Q 6Plex no PC
5refurbished 9001960 Rotor-Gene AusDx (Outback Red) 9002000 Rotor-Gene Q MDx 2Plex 9002001 Rotor-Gene Q MDx 2Plex no PC 9002010 Rotor-Gene Q MDx 2Plex HRM 9002011 #Rotor-Gene Q MDx 2Plex HRM no PC 9002020 Rotor-Gene Q MDx 5Plex 9002021 Rotor-Gene Q MDx 5Plex no PC 9002029 Rotor-Gene Q MDx 5plex, PrioPLUS 9002030 Rotor-Gene Q MDx 5Plex HRM 9002030R Rotor-Gene Q MDx 5Plex HRM
6refurbished 9002031 Rotor-Gene Q MDx 5Plex HRM no PC 9002040 Rotor-Gene Q MDx 6Plex 9002041 Rotor-Gene Q MDx 6Plex no PC90275010 9002204 Rotor-Gene Q DBV 9023241 Rotor-Gene Q Software 9002370 5 plex MDx HRM with laptop (canada) This issue can affect the following QIAGEN IVD kits: Catalogue number Product Name 4514363 artus BK Virus QS-RGQ Kit (24) 4514263 artus BK Virus RG PCR Kit (24) 4514265 artus BK Virus RG PCR Kit (96) 4503363 artus CMV QS-RGQ Kit (24) 4503263 artus CMV RG PCR Kit (24) 4503265 artus CMV RG PCR Kit (96) 4501363 artus EBV QS-RGQ Kit (24) 4501263 artus EBV RG PCR Kit (24) 4501265 artus EBV RG PCR Kit (96) 4506366 artus HBV QS-RGQ (72) 4506356 artus HBV QS-RGQ Kit (24) 4506363 artus HBV QS-RGQ Kit (24) 4506263 artus HBV RG PCR Kit (24) 4506265 artus HBV RG PCR Kit (96) 4518356 artus HCV QS-RGQ Kit (24) 4518363 artus HCV QS-RGQ Kit (24) 4518366 artus HCV QS-RGQ Kit (72) 4538366 artus HCV QS-RGQ Kit V2 4518253 artus HCV RG RT-PCR Kit (24) 4518263 artus HCV RG RT-PCR Kit (24) 4518265 artus HCV RG RT-PCR Kit (96) 4513363 artus HI Virus-1 QS-RGQ Kit (24) 4513366 artus HI Virus-1 QS-RGQ Kit (72) 4513253 artus HI Virus-1 RG RT-PCR Kit (24) 4513263 artus HI Virus-1 RG RT-PCR Kit (24) 4513265 artus HI Virus-1 RG RT-PCR Kit (96) 4502363 artus VZV QS-RGQ Kit (24) 4502263 artus VZV RG PCR Kit (24) 4502265 artus VZV RG PCR Kit (96) 4532265 artus JCV RG PCR Kit (96) 4555263 artus M.tuberculosis RG PCR Kit (24) 4555265 artus M.tuberculosis RG PCR Kit (96) 4504263 artus Parvo B19 RG PCR Kit (24) 4511263 artus SARS RG RT-PCR Kit (24)
Amount recalled
11418 instruments

Where it was sold

  • Nationwide
  • GA

Why it was recalled

When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the calculated concentration result value as a logarithmic value, and could lead to a false negative result, which could lead to serious medical consequences such as the suspension or non-initiation of treatment

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Qiagen Sciences LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 26, 2024 · Data as of Sep 19, 4:49 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.