Reported Apr 10, 2024 · Ongoing

stryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplasty

Wright Medical Technology, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Wright Medical Technology, Inc. · 1023 Cherry Rd · Memphis, TN · 38117-5423

Is this the product you have?

  • stryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplasty
Lot codes and dates
  • UDI/DI 00889797004008
  • Lot Numbers
    • 2656950
    • 2762126
    • 2796094
Amount recalled
47 units

Where it was sold

Why it was recalled

Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Wright Medical Technology, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:59 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.