Reported Mar 25, 2020 · Terminated
UroDiagnost Eleva, system code 708033
Philips North America, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips North America, LLC · 3000 Minuteman Rd · Andover, MA · 01810-1032
Is this the product you have?
- UroDiagnost Eleva, system code 708033
- Lot codes and dates
- Serial numbers 64 48 2 60
- Amount recalled
- 4
This recall covers 10 products
The agency filed one notice per product. You are reading one of them; the other 9 are below.
Where it was sold
- Nationwide
Why it was recalled
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.