Reported Mar 29, 2017 · Terminated
Giraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a…
Ohmeda Medical
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ohmeda Medical · 8880 Gorman Rd · Laurel, MD · 20723-5800
Is this the product you have?
- Giraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SP02 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.
- Lot codes and dates
- See Attached Spreadsheat
- Amount recalled
- 5,993 (5,080 US; 913 OUS)
This recall covers 11 products
The agency filed one notice per product. You are reading one of them; the other 10 are below.
Where it was sold
- Nationwide
- PR
Why it was recalled
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Ohmeda Medical recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.