Reported Aug 17, 2022 · Ongoing

Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High…

GE Healthcare, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

  • Service manuals for the CT, PET
  • or NM table as followsGT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite Table, Kunlun Table, Global PET-CT Table, NP Table, P2005AA Table, LCT Table, CTE TWIN TABLE Assy. All devices are patient tables for PET/CT systems
Lot codes and dates

49 of 49 products

Products covered by this recall
Model / numberBarcodesProduct as published
(1) Service manual identifier 5193754-800, Rev 23 and prior, GTIN: N/A
00840682102490(2) Service manual identifier 5307450-2EN, Rev 30 and prior, GTIN: Revolution Discovery CT: 00840682102490
00840682125574Revolution Discovery CT A: 00840682125574
Revolution HD: N/A
Discovery CT750 HD: N/A
00840682102551Discovery CT: N/A. NOTE: Revolution Discovery CT has two different GTINs to identify different versions of the device. (3) Service manual identifier 5350500-8EN, Rev 28 and prior, GTIN: Optima CT540: 00840682102551
00195278563408.00840682138963001952785634080084068211651000840682146180Optima CT540 N: 00840682138963. Note: Optima CT540 has two different GTINs to identify different versions of the device. (4) Service manual identifier 5836560-8EN, Rev 2 and prior, (5) Service manual identifier 5366080-8EN, Rev 39 and prior, GTIN: Optima CT660 Cj2.5 Upgrade: 00840682116510. (6) Service manual identifier 5487410-8EN, Rev 14 and prior, GTIN: N/A. (7) Service manual identifier 5809950-8EN, Rev 3 and prior, GTIN: Revolution Maxima: 00840682146180
00840682147392Revolution Maxima US: 00840682147392
00840682146159Revolution Ace: 00840682146159
Optima CT680 Expert: 008406821469372
00840682146173.00195278420893008406821375530084068211869900840682105286008406821097960084068214617300840682102490Revolution Maxima M: N/A. (8) Service manual identifier 5848536-8EN, Rev 1 and prior, GTIN: 00195278420893. (9) Service manual identifier 5796213-2EN, Rev 8 and prior, GTIN: 00840682137553. (10) Service manual identifier 5366638-8EN, Rev 24 and prior, GTIN: Discovery RT: 00840682118699, Discovery CT590RT and CT580: N/A. (11) Service manual identifier 5196837-800, Rev 12 and prior, GTIN: N/A. (12) Service manual identifier 5193574-800, Rev 36 and prior, GTIN: 00840682105286. (13) Service manual identifier 5193574-800, Rev 36 and prior, GTIN: N/A. (14) Service manual identifier 5866660-8EN, Rev 26 and prior, GTIN: 00840682109796. (15) Service manual identifier 5987660-8EN, Rev3 and prior, (16) Service manual identifier 5307450-2EN, Rev 30 and prior, GTIN: Revolution Discovery CT: 00840682102490
00840682125574Revolution Discovery CT A: 00840682125574
Revolution HD: N/A
Discovery CT750 HD: N/A
0084068211651000840682146180Discovery CT: N/A. NOTE: Revolution Discovery CT has two different GTINs to identify different versions of the device. (17) Service manual identifier 5366080-8EN, Rev 39 and prior, GTIN: Optima CT660 Cj2.5 Upgrade: 00840682116510. (18) Service manual identifier 5809950-8EN, Rev 3 and prior, GTIN: Revolution Maxima: 00840682146180
00840682147392Revolution Maxima US: 00840682147392
00840682146159Revolution Ace: 00840682146159
Optima CT680 Expert: 008406821469372
00840682137553.0084068213755300840682143523Revolution Maxima M: Not applicable. NOTE: Revolution Maxima has two different GTINs to identify different versions of the device. (19) Service manual identifier 5796213-2EN, Rev 8 and prior, (20) Service manual identifier 5196837-800, Rev 12 and prior, GTIN: N/A. (21) Service manual identifier 5661743-2EN, Rev 9 and prior, GTIN: Discovery MI weldment: 00840682143523
00840682108218Discovery MI non-weldment: 00840682108218
Amount recalled
24,843 tables

Where it was sold

Why it was recalled

As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Healthcare, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:58 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.