Reported May 5, 2021 · Ongoing
Philips Xper Flex Cardio (FC2010, Procedure Room and FC2020, Bedside Monitor) devices - intended for use by…
Philips North America LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips North America Llc · 222 Jacobs St · Cambridge, MA · 02141-2289
Is this the product you have?
- Philips Xper Flex Cardio (FC2010, Procedure Room and FC2020, Bedside Monitor) devices - intended for use by professional healthcare providers for complete physiologic/hemodynamic monitoring
- Service numbers associated with FC2010Xper Flex Cardio 2010 Rev A 453564241901 Xper Flex Cardio 2010 Rev B 453564483321 Flex Cardio FC2010 Rev C 453564621791 Xper Flex Cardio 2010 REV C RUEXCHANGE 453564675021 Xper Flex Cardio 2010 Rev D 453564669081 Xper Flex Cardio 2010 Rev D Russia 453564674581 Xper Flex Cardio FC2010 Rev E 453564845841 Xper Flex Cardio 2010 Rev E Russia 453564845881
- Service numbers associated with FC2020Xper Flex Cardio 2020 Rev A 453564241911 Xper Flex Cardio 2020 Rev B 453564483331 Flex Cardio FC2020 Rev C 453564621801 Xper Flex Cardio 2020 Rev D 453564675001 Xper Flex Cardio FC2020 Rev
- E 453564845861 Software version number1.5.19 and 1.6.8
- Lot codes and dates
- Service Number/ UDI # 453564241901 (01)00884838082113; 453564241911 (01)00884838082113; 453564483321 (01)00884838082113; 453564483331 (01)00884838083516; 453564621791 (01)00884838084902; 453564621801 (01)00884838084919; 453564845881 (01)00884838095052; 453564675021 (01)00884838103245; 453564669081 (01)00884838083820; 453564675001 (01)00884838086005; 453564674581 (01)00884838083820; 453564845841 (01)00884838093645; 453564845861 (01)00884838093652
- Amount recalled
- 9208 units WW
Where it was sold
- Nationwide
- PR
Why it was recalled
Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the nearest whole kilogram;Xper IM software used with the Xper Flex Cardio Physio Monitoring System may periodically crash; No SpO2 numeric or plethysmography is displayed when SpO2 is connected o the Flex Cardio device; display of certain data from the FC2010 becomes frozen, i.e., waveforms cease sweeping and updating and the ECG, IBP, and respiration numeric values cease to update;e ECG, IBP, and respiration waveforms become flat lines and no audible alarms are produced for HR and IBP, upon start up, an unexpected non physiological ECG waveform, erratic heart rate numeric value, and non-physiological display of any other active waveforms may appear on the Boom Monitor
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new produce recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.