Reported May 2, 2018 · Terminated
Quidel MicroVue Intact PTH EIA, Model 8044. An enzyme immunoassay for the quantification of Intact Parathyroid Hormone…
Diagnostic Hybrids, Inc.
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Diagnostic Hybrids, Inc. · 2005 E State St Ste 100 · Athens, OH · 45701-2125
Is this the product you have?
- Quidel MicroVue Intact PTH EIA, Model 8044. An enzyme immunoassay for the quantification of Intact Parathyroid Hormone (PTH) in serum.
- Lot codes and dates
- Lot 068259
- Amount recalled
- 26
This recall covers 5 products
The agency filed one notice per product. You are reading one of them; the other 4 are below.
Why it was recalled
There may be a loss of separation between Calibrators A and B which may impact the assay calibration at the lower end of the assay.