Reported Jun 19, 2013 · Terminated
Applied Medicals Epix¿ and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for grasping and…
Applied Medical Resources Corp.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Applied Medical Resources Corp · 22872 Avenida Empresa # 3 · Rancho Santa Margarita, CA · 92688-2650
Is this the product you have?
- Applied Medicals Epix¿ and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for grasping and manipulating tissue during general or laparoscopic surgery.
- Lot codes and dates
- Model Numbers C4130 and C4140. MODEL NUMBERS C4130- 5mm x 35cm Epix/Direct Drive Laparoscopic Grasper LOT NUMBERS 1145332, 1145943, 1145944, 1145945, 1145992, 1145993, 1145994, 1146371, 1147900, 1147901, 1147955, 1148172, 1148173, 1148174, 1148180, 1148181, 1148184, 1148190, 1148191, 1148192, 1148194, 1148201, 1148203, 1148677, 1148765, 1148766, 1148767, 1148835, 1149040, 1149041, 1149690, 1149757, 1150168, 1150169, 1150170, 1150490, 1150634, 1151246, 1151247, 1151248, 1152274, 1152728, 1153332, 1154394, 1154637, 1155135, 1155473, 1155948, 1156004, 1156010, 1156011, 1156012, 1156270, 1156360, 1156452, 1156577, 1156857, 1157218, 1157679, 1157950, 1157951, 1157952, 1157953, 1157954, 1158052, 1158131, 1158216, 1158217, 1158425, 1159491, 1160167, 1161136, 1161137, 1161398, 1161638, 1162112, 1162503, 1162504, 1162588, 1163399, 1163558, 1163559, 1163575, 1164580, 1164581, 1164582, 1164583, 1168124, 1168637, 1168638, 1169117, 1171168, 1171631, 1172170, 1172668, 1173153, 1174995, 1174996, 1176616, 1177407, 1177408, 1178109, 1180295, 1181554, 1182940, 1183774, 1184416, 1184959, 1187241, 1188110, 1188745, 1189773, 1191368, 1191809. C4140- 5mm x 45cm Epix/Direct Drive Laparoscopic Grasper 1145559, 1147851, 1147852, 1147960, 1148210, 1148211, 1149038, 1149039, 1149689, 1149936, 1150488, 1150489, 1151249, 1152594, 1153330, 1155645, 1157217, 1157949, 1158644, 1160033, 1160170, 1162111, 1167871, 1169855, 1171497, 1172009, 1173587, 1174121, 1177578, 1179347, 1180753, 1181346, 1182924, 1183769, 1184961, 1188111, 1189182 Various Kits Containing Either C4130 or C4140 1148079, 1148139, 1148145, 1148250, 1148514, 1148770, 1148780, 1149100, 1149605, 1150455, 1151831, 1153074, 1153702, 1154931, 1154997, 1155332, 1156581, 1156812, 1157090, 1158535, 1160152, 1160227, 1160388, 1161766, 1162189, 1162343, 1163348, 1163881, 1164590, 1164796, 1164831, 1165443, 1166520, 1168227, 1169190, 1169456, 1169620, 1169772, 1170151, 1170378, 1171280, 1171878, 1171879, 1172330, 1173319, 1173903, 1174093, 1175987, 1175991, 1176171, 1178210, 1178498, 1178581, 1179376, 1182679, 1182972, 1183303, 1184673, 1185599, 1185762, 1186257, 1186868, 1187082, 1188827, 1189581, 1190574, 1191055, 1193112, 1193147, 1193648, 1193688, 1194964, 1194965
- Amount recalled
- 127,305 units
Where it was sold
- Nationwide
Why it was recalled
Applied Medical is conducting a voluntary recall on specific lot numbers of its Epix¿ and Direct Drive Laparoscopic Graspers. When the ratchet trigger is activated, if a great enough force is applied, there is a potential for the trigger to fracture and the jaws to remain in the closed position. The likelihood of this situation to occur and result in permanent patient injury is highly unlikely;
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Applied Medical Resources Corp. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.