Reported May 12, 2021 · Ongoing
Programmers - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also…
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · 8200 Coral Sea St Ne · Mounds View, MN · 55112-4391
Is this the product you have?
- Programmers - Product Usageused to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. 2090 Carelink 2. 29901 Encore
- Lot codes and dates
- all codes
- Amount recalled
- 37690 units
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Why it was recalled
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medtronic Inc., Cardiac Rhythm and Heart… recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.