Reported Jun 19, 2013 · Terminated
embla Embletta Gold (It functions as a Polysomnography Recorder and Ventilator Effort Recorder). The Embletta Gold is a…
Embla System LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Embla System Llc · 250 Cooper Ave · Tonawanda, NY · 14150-6633
Is this the product you have?
- embla Embletta Gold (It functions as a Polysomnography Recorder and Ventilator Effort Recorder). The Embletta Gold is a battery operated device which records physiologic signals used in the diagnosis of sleep disorders. It is intended to be used for adult and pediatric (excluding neonatal and infant) studies. The recorder is not equiped with an alarm device and is not intended to be used as a life monitor. It may be used in a variety of configurations and the recording of physiologic signals of a sleeping patient
- Item NumbersUI_000670 UI_001339
224 of 224 codes
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
UI004904RMIV007125UI014813UI017060UI019513RMIV019041144913RMIV019639RMIV0197781010002678101000432510100043101596991010003947UI011502UI020772
- Lot codes and dates
- Serial Numbers
516 of 516 codes
EBG-C1000497C0914092-EBGEBG-C1002589EBG-C1002598EBG-C1001659EBG-C1001832EBG-C1001835EBG-C1001843EBG-C1001845EBG-C1002253EBG-C1002473EBG-C1002504EBG-C1002519EBG-C1002537EBG-C1002539EBG-C1001367EBG-C1001378EBG-C1001763EBG-C1000265EBG-C1002607EBG-C1002542EBG-C1003212EBG-C1001834EBG-C1001858EBG-C1001889EBG-C1002445EBG-C1003052EBG-C1002502EBG-C1002508EBG-C1001776C0901065-EBGEBG-C1003078EBG-C1003084EBG-C1003085EBG-C1003161EBG-C1003165EBG-C1001230EBG-C1001275EBG-C1002688EBG-C1002500EBG-C1003055EBG-C1001970EBG-C1001972EBG-C1001973EBG-C1001983EBG-C1001986EBG-C1001518EBG-C1000112EBG-C1001308EBG-C1002771EBG-C1002863EBG-c1001330EBG-C1001237C0901015-EBGEBG-C1000118EBG-C1001394EBG-C1001411EBG-C1002871EBG-C1003067EBG-C1002602
- C0914086-EBG EBG-C1001368 Item NumbersUI_000670 UI_001339
224 of 224 codes
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
UI004904RMIV007125UI014813UI017060UI019513RMIV019041144913RMIV019639RMIV0197781010002678101000432510100043101596991010003947UI011502UI020772
- Serial Numbers
- Amount recalled
- 519 units were calibrated at the service center
Where it was sold
Why it was recalled
Calibration error for Embla Embletta Gold units calibrated through service center between June 15, 2012 to February 27, 2013. This calibration error could cause a clinician to over titrate the patient and prescribe a therapy pressure higher than is necessary.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new infant & baby recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.