Reported Apr 1, 2020 · Terminated
FlipCutter III Drill
T.a.g. Medical Products Corporation, Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: T.A.G. MEDICAL PRODUCTS CORPORATION, LTD. · Kibbutz Gatton · Naharia, N/A · N/A
Is this the product you have?
- FlipCutter III Drill
- Lot codes and dates
- UDI10818674021446
- Part NumberAR-1204FF
- Lots
79 of 79 codes
19E0119E0219E0319E0519E1019E1119F0119F0219F0319F0419F0519F0619F0719F0819F0919F1019J1519J1619J1719J1819J1919J2019J2119J2219J2319J2419J2519J2619J2819J2919K0119K0219K0319K0919K1119K1219M0119M0919M1019M1119M1219M1319M1419M1519M1619M1719M1819M2119M2219N0219N0319N0519N0619N0719N0819N0919N1119N1219N1319N14
- Amount recalled
- 6625 units
Why it was recalled
There is potential for the blister tray to be cracked causing the sterile barrier to be compromised and potential for particulate within the sterile barrier.
This product could cause temporary or reversible health problems — stop using it.