Reported May 4, 2016 · Terminated
3M Steri-Vac Gas Sterilizer/Aerator, models 5XL and 8XL.
3M Company - Health Care Business
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: 3M Company - Health Care Business · 3m Center 2510 Conway Ave , B# 275-5-W-6 · Saint Paul, MN · 55144-0001
Is this the product you have?
3M Steri-Vac Gas Sterilizer/Aeratormodels 5XL and 8XL
- Lot codes and dates
- Model 5XL
- Serial Numbers
722612 - 722683820352 - 820360 Model 8XL
- Serial Numbers
351303 - 351399450144 - 450147
- Amount recalled
- 74 valves
Where it was sold
Why it was recalled
3M discovered a valve with incorrectly labeled "input" and "output" orientation. This defect can result in the valve being installed backwards, creating a remote possibility of valve failure. A failure may result in EO being vented outside the sealed chamber during a limited period in the EO injection phase of the cycle.