Reported May 19, 2021 · Terminated

Calibration Serum Level 3 (CAL 3) - multi-analyte mixture calibrator. Catalogue Number: CAL2351

Randox Laboratories Ltd.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Randox Laboratories Ltd. · 55 Diamond Road · Crumlin (North) · N/A

Is this the product you have?

  • Calibration Serum Level 3 (CAL 3) - multi-analyte mixture calibrator
  • Catalogue NumberCAL2351
Lot codes and dates
  • Batch / Lot number1162UE
  • GTIN05055273200966
Amount recalled
98 kits

Where it was sold

Why it was recalled

Randox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator CAL2351, lot 1162UE. The positive bias has been attributed to bilirubin instability in the calibrator and is evident in both Quality Control (QC) and Patient samples.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Randox Laboratories Ltd. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 22, 2021 · Data as of Sep 19, 5:43 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.