Reported May 4, 2016 · Terminated
Automated peritoneal dialysis (APD) cycler
Baxter Healthcare Corp.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Baxter Healthcare Corp. · 1 Baxter Pkwy · Deerfield, IL · 60015-4625
Is this the product you have?
- Automated peritoneal dialysis (APD) cycler
- Lot codes and dates
- N/A
- Amount recalled
- 10 units
Where it was sold
Why it was recalled
Baxter Healthcare Corporation is sending this communication to inform you of incomplete instructions in the AMIA Automated Peritoneal Dialysis (PO) System Clinician Guide for calculating the total recommended solution therapy volume. Specifically, the instructions do not specify the need for an extra 200 mL of PD solution in order to prime the patient line and for air purge operations. The tot
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Baxter Healthcare Corp. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.