Reported May 4, 2016 · Terminated

enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V or 110V -- COMMON/USUAL…

Ortho-Clinical Diagnostics

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ortho-Clinical Diagnostics · 100 Indigo Creek Dr · Rochester, NY · 14626-5101

Is this the product you have?

  • enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V
  • or 110V -- COMMON/USUAL NAMEenGen (TM)
  • Track System -- NOTEOutside the US the system is also known as VITROS Automation Solution (VAS) TC Automation System (configured with Thermo Scientific Recapper Module from VITROS Automation Solutions), Product Code 6844300 --- The enGen (TM) Lab Automation System is a modification to VITROS Analyzers that includes the addition of an automated track sub-system (TC Automation) for pre-analytical and post-analytical sample and data management. In addition to an Automated Track Sub-system, the enGen (TM) System includes one or more VITROS Chemistry, Immunodiagnostic and/or Integrated Systems (clinical analyzers) as well as a Data Management Subsystem (Middleware), and a Software Configuration (gsb) File. The enGen System is typically interfaced with a Laboratory Information System (LIS) via connectivity facilitated at the Data Management Subsystem level. The enGen (TM) Laboratory Automation System contains a module (i.e. enGen (TM)Laboratory Automation Recapper Module) that applies caps to the sample collection container after the sample has been aspirated. In this module, an electrical cable supplies power to the cap vibratory feeder and is supposed to be located above the vibratory feeder support plate. The cap vibratory feeder sits upon a feeder support plate. Un-capped sample containers pass underneath this support plate prior to being re-capped for storage. The Recapper Module is supplied by Thermo Fisher Scientific
Lot codes and dates
  • US Serial Numbers
    • JFBX395J
    • GBX395J
    • J953252
    • JJXBQW4J
    • J3RHC23J
    • J86PWC5J
    • J5LQ6Q2J
    • JFCM073J
    • J54FC01J
    • J5J4B45J
    • J7KSC15J
    • J4HX7S4J
    • JGWXQ52J
    • J953221
    • JG7BFB5J
    • JHPKG25J
    • JCXL941J
    • J736BM4J
    • JFXCBG3J
    • --
  • Foreign Serial Numbers
    • J89133 (VAS)
    • JHWJZY4J
    • J3J4B45J
    • J953244
    • JCT20Q3J
    • J16XQH4J
    • GWJVY4J
    • DWX052J
    • J271264
    • J271368
    • J271371
    • J271269
    • J85206
    • J953216
    • DP2LF62
    • J7023874
    • J7023847
    • J7025017
    • J7024893
    • J7023645
    • J7023435
    • J57663
    • J800103
    • S/N 89VQ762 (VAS)
    • J75863
    • J75691
    • J30372
    • J3055503
    • J3055496
Amount recalled
US: 19, Foreign: 30

Where it was sold

Why it was recalled

The electrical cable that supplies power to the cap vibratory feeder of the Recapper Module may loosen from its correct position above the vibration feeder support plate and migrate to under the support plate. If this occurs, the uncapped sample tubes may hit the cable, and sample fluid (e.g., serum) may splash from the tube and into another tube or onto the under carriage of the support plate.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ortho-Clinical Diagnostics recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 30, 2018 · Data as of Sep 19, 5:21 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.