Reported Mar 25, 2026 · Ongoing

MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Howmedica Osteonics Corp.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Howmedica Osteonics Corp. · 325 Corporate Dr · Mahwah, NJ · 07430-2006

Is this the product you have?

  • MICS3 Angled Sagittal Saw Attachment
  • Part Number210490
Lot codes and dates
  • Part Number210490
  • UDI(01)07613327523874(10)TU6563205RP
  • Lot numbersALL
Amount recalled
1098 units

Where it was sold

  • Nationwide

Why it was recalled

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Howmedica Osteonics Corp. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:52 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.