Reported Apr 24, 2024 · Ongoing

Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426

Randox Laboratories Ltd.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Randox Laboratories Ltd. · 55 Diamond Road · Crumlin (North) · N/A

Is this the product you have?

  • Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate
  • the Microalbumin Assay Catalog NumberMA2426
Lot codes and dates
  • GTIN05055273204544 (1)
  • Batch/Lot602114 /1083MA-1087MA (2)
  • Batch/Lot636645 /1108MA-1112MA (3)
  • Batch/Lot662429 /1083MA- 1087MA
Amount recalled
6 units

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Randox Laboratories Ltd. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:40 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.