Reported Apr 8, 2020 · Terminated

Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000

Stryker Communications

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Stryker Communications · 571 Silveron Blvd · Flower Mound, TX · 75028

Is this the product you have?

  • Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Lot codes and dates
  • Serial Number Range091316 70016 through 083117-73252
Amount recalled
1415 Surgical Lights

Where it was sold

  • Nationwide

Why it was recalled

The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Stryker Communications recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 18, 2022 · Data as of Sep 19, 6:11 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.