Reported Apr 8, 2020 · Terminated
Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Stryker Communications
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Stryker Communications · 571 Silveron Blvd · Flower Mound, TX · 75028
Is this the product you have?
- Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
- Lot codes and dates
- Serial Number Range091316 70016 through 083117-73252
- Amount recalled
- 1415 Surgical Lights
Where it was sold
- Nationwide
Why it was recalled
The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.
This product could cause temporary or reversible health problems — stop using it.