Reported May 1, 2024 · Ongoing
LightStrike Blackout Curtains, Part numbers; 610-00008, 610-00009, 610-00019, 610-00020, and 610-00024, used with the…
XENEX Disinfection Services Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: XENEX Disinfection Services Inc. · 1074 Arion Cir Ste 116 · San Antonio, TX · 78216-3085
Is this the product you have?
LightStrike Blackout CurtainsPart numbers610-00008610-00009610-00019610-00020and 610-00024used with the
- LightStrike Germ-Zapping Robots Models
PXUV4DPXUV5DLSMUV6-FTor LSMUV6-SLfor whole room microbial reduction
- Lot codes and dates
- Robot UDI-DIs
00814122020033008141220200640081412202012500814122020132. All serial numbers
- Robot UDI-DIs
- Amount recalled
- 377
Where it was sold
Why it was recalled
Blackout curtains are no longer recommended as a protective barrier against UV/Ozone exposure to a bystander standing in the same room, but on the other side of the curtain from where a microbial reduction device UV cycle is taking place. Bystanders in the room during the cycle may experience temporary skin reddening, feeling of sand in eyes, nausea, headache, and/or mucous membrane irritation.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.