Reported May 31, 2023 · Ongoing
Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number:…
OmniCell, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: OmniCell, Inc. · 51 Pennwood Pl Ste 400 · Warrendale, PA · 15086-6506
Is this the product you have?
- Omnicell IVX Station-For use in the automated preparation of pharmaceutical
- admixtures and syringes Model Number298921373
- Lot codes and dates
- UDI-DI00860006285005
- Serial Numbers
IVXS000014IVXS000015IVXS000016IVXS000017IVXS000018. IVXS000019
- Amount recalled
- 6 units
Where it was sold
Why it was recalled
IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling