Reported Apr 8, 2020 · Ongoing
Philips V60 Ventilators with Power Management PCBA part number 1055906
Respironics California, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Respironics California, LLC · 2271 Cosmos Ct · Carlsbad, CA · 92011-1517
Is this the product you have?
- Philips V60 Ventilators with Power Management PCBA part number 1055906
- Lot codes and dates
- Serial Numbers
100002908 to 100027093100027096 to 100085389100085399 to 100109746100109868 to 201003390201003392 to 201010952
- Serial Numbers
- Amount recalled
- 23581
Where it was sold
Why it was recalled
Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Respironics California, LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.