Reported May 7, 2025 · Ongoing
GE Healthcare Carestation 620 SE, Model/REF Number 1012-9620-012. Anesthesia system.
GE Medical Systems China Co., Ltd.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Medical Systems China Co., Ltd. · Dev. Zone · Wuxi · N/A
Is this the product you have?
- GE Healthcare Carestation 620 SE, Model/REF Number 1012-9620-012. Anesthesia system.
- Lot codes and dates
- GTIN 00195278569677
- Serial Numbers
SM624250018WCSM624250019WCSM624500025WCSM624500026WCSM624500027WCSM624290020WCSM624290021WCSM623470001WCSM623470002WCSM623470003WCSM623470004WCSM623470005WCSM623470006WCSM623470007WCSM623470008WCSM623470009WCSM623470010WCSM623470011WCSM623470012WCSM624240017WCSM624240013WCSM624240014WCSM624240015WCSM624240016WCSM624110001WCSM624330022WCSM624330023WC
- Amount recalled
- 27 units
This recall covers 15 products
The agency filed one notice per product. You are reading one of them; the other 14 are below.
2 more are in the archive.
Where it was sold
- Nationwide
- PR
Why it was recalled
Stop using this product now
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new GE Medical Systems China Co., Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.