Reported May 16, 2018 · Terminated

ARCHITECT DHEA-S Reagent Product Usage: It is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative…

Abbott GmbH & Co. KG

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Abbott Gmbh & Co. KG · Max-Planck-Ring 2 · Wiesbaden, N/A · N/A

Is this the product you have?

  • ARCHITECT DHEA-S Reagent Product UsageIt is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of dehydroepiandrosterone sulfate (DHEA-S) in human serum and plasma on the ARCHITECT iSystem
Lot codes and dates

11 of 11 products

Products covered by this recall
#BarcodesProduct as published
100380740017446Lot No. 01316K000, Exp. Date 04FEB2018, UDI (01) 00380740017446 (17) 180204 (10) 01316K000 (240) 8K2720
200380740017446Lot No. 01217A000, Exp. Date 13APR2018, UDI (01) 00380740017446 (17) 180413 (10) 01217A000 (240) 8K2720
300380740017446Lot No. 02217D000, Exp. Date 24JUN2018, UDI (01) 00380740017446 (17) 180624 (10) 02217D000 (240) 8K2720
400380740017446Lot No. 04117H000, Exp. Date 25NOV2018, UDI (01) 00380740017446 (17) 181125 (10) 04117H000 (240) 8K2720; List No. 8K27-25 with the following Lot No., Exp. Date, UDI
500380740017453Lot No. 01216K000, Exp. Date 04FEB2018, UDI (01) 00380740017453 (17) 180204 (10) 01216K000 (240) 8K2725
600380740017453Lot No. 01117A000, Exp. Date 13APR2018, UDI (01) 00380740017453 (17) 180413 (10) 01117A000 (240) 8K2725
700380740017453Lot No. 02217C000, Exp. Date 17JUN2018, UDI (01) 00380740017453 (17) 180617 (10) 02217C000 (240) 8K2725
800380740017453Lot No. 02317D000, Exp. Date 24JUN2018, UDI (01) 00380740017453 (17) 180624 (10) 02317D000 (240) 8K2725
900380740017453Lot No. 02417D000, Exp. Date 24JUN2018, UDI (01) 00380740017453 (17) 180624 (10) 02417D000 (240) 8K2725
1000380740017453Lot No. 00517F000, Exp. Date 02SEP2018, UDI (01) 00380740017453 (17) 180902 (10) 00517F000 (240) 8K2725
1100380740017453Lot No. 02817H000, Exp. Date 21OCT2018, UDI (01) 00380740017453 (17) 181021 (10) 02817H000 (240) 8K2725
Amount recalled
10,153 units

Where it was sold

Why it was recalled

Falsely elevated results may be obtained when using the ARCHITECT DHEA-S assay with samples from infants up to 60 days old.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new infant & baby recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 10, 2018 · Data as of Sep 19, 5:26 PM UTC