Reported Jun 2, 2021 · Terminated
BiPAP A40, Brazil: Continuous ventilator, non-life-supporting. Material Number: BR1111169
Philips Respironics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Respironics, Inc. · 1001 Murry Ridge Ln · Murrysville, PA · 15668-8517
Is this the product you have?
- BiPAP A40
- BrazilContinuous ventilator, non-life-supporting
- Material NumberBR1111169
- Lot codes and dates
- Serial Number
V306120599B37V3061207124CFV30612105753FV306121267414
- Serial Number
- Amount recalled
- 4 units
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
Why it was recalled
Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Respironics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.