Reported Sep 7, 2022 · Ongoing
ARCHITECT STAT Troponin-I Reagent Kit, REF 2K41-27
Abbott Laboratories
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Abbott Laboratories · 100 Abbott Park Rd Bldg Ap8b · Abbott Park, IL · 60064-3502
Is this the product you have?
- ARCHITECT STAT Troponin-I Reagent Kit, REF 2K41-27
- Lot codes and dates
- UDI (01) 00380740048525 (17) 991231 (10) 12345M100 (240) 2K4127 Lot lot 63718UN22
- Amount recalled
- 2451 kits of 100 tests
Where it was sold
- Nationwide
Why it was recalled
One lot of reagent was manufactured with insufficient quantities of the heterophilic blocking agents in the conjugate component.