Reported Jun 7, 2023 · Ongoing

Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that…

Teleflex LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: TELEFLEX LLC · 3015 Carrington Mill Blvd · Morrisville, NC · 27560-5437

Is this the product you have?

  • Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or tissue ligation
Lot codes and dates
  • UDI/DI 24026704696394
  • Lot Numbers
    • 73A2200857
    • 73A2300018
    • 73A2300176
    • 73A2300326
    • 73A2300520
    • 73A2300778
    • 73D2200216
    • 73D2200559
    • 73E2200634
    • 73E2200874
    • 73E2201099
    • 73F2200193
    • 73F2200200
    • 73F2200482
    • 73F2200711
    • 73G2200168
    • 73G2200345
    • 73G2200560
    • 73G2200754
    • 73H2200115
    • 73H2200294
    • 73H2200433
    • 73H2200453
    • 73J2200747
    • 73J2200763
    • 73K2200014
    • 73K2200095
    • 73K2200101
    • 73K2200299
    • 73K2200566
    • 73K2200713
    • 73L2200064
    • 73L2200066
Amount recalled
146,606 units

Why it was recalled

Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Teleflex LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:14 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.