Reported Apr 30, 2025 · Ongoing

The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall…

Abbott Molecular, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Abbott Molecular, Inc. · 1300 E Touhy Ave · Des Plaines, IL · 60018-3315

Is this the product you have?

  • The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).
Lot codes and dates
  • All Serial Numbers. 08N53-001
  • UDI-DI
    • 00884999047389
    • 08N53-002
  • UDI-DI
    • 00884999048034
    • 08N53-032
  • UDI-DI00884999047587
Amount recalled
N/A

Where it was sold

  • Nationwide

Why it was recalled

The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Abbott Molecular, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:38 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.