Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number…
Mindray DS USA, Inc. dba Mindray North America
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Mindray DS USA, Inc. dba Mindray North America · 800 MacArthur Blvd · Mahwah, NJ · 07430-2001
Is this the product you have?
- Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.
- Lot codes and dates
- Part No0998-00-0708-01 Software version 8.9
- Amount recalled
- 4 units
Where it was sold
- Nationwide
- TX
Why it was recalled
Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Mindray DS USA, Inc. dba Mindray North America recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.