Reported Jun 3, 2015 · Terminated

MEDRAD¿ Stellant CT Injector System with Certegra Workstation; intended for the specific purpose of injecting…

Medrad Inc. dba Bayer R&I

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Medrad Inc dba Bayer R&I · 1 Medrad Dr · Indianola, PA · 15051-9759

Is this the product you have?

  • MEDRAD¿ Stellant CT Injector System with Certegra Workstation; intended for the specific purpose of injecting intravenous contrast media into humans for diagnostic studies in computed tomography (CT) applications.
Lot codes and dates
  • Material numbers
    • 3032458
    • 3007301
    • 3010432
  • with serial numbers

    137 of 137 codes

    • 10054
    • 37918
    • 37921
    • 36714
    • 36721
    • 108
    • 40596
    • 23834
    • 10065
    • 39533
    • 20092
    • 29884
    • 33423
    • 10026
    • 10028
    • 10032
    • 10035
    • 10069
    • 29881
    • 30351
    • 23741
    • 27877
    • 10095
    • 20693
    • 30548
    • 33316
    • 21089
    • 10123
    • 10008
    • 24272
    • 37177
    • 25989
    • 39003
    • 39009
    • 38894
    • 38894
    • 38840
    • 38843
    • 34894
    • 38833
    • 35824
    • 38846
    • 26107
    • 37073
    • 37472
    • 10058
    • 10010
    • 10076
    • 24707
    • 10091
    • 38400
    • 38403
    • 38404
    • 38405
    • 40390
    • 10093
    • 38999
    • 39434
    • 37351
    • 37359
Amount recalled
133

Where it was sold

  • Nationwide

Why it was recalled

The root cause investigation showed that the under-volume hazard can occur when the unit has been configured to prime with contrast, but instead primes with saline due to the configuration data not being sent to the injector head. This may create a non-optimal priming condition that could lead to under-infusion.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Medrad Inc. dba Bayer R&I recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 16, 2015 · Data as of Sep 19, 5:34 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.