Reported May 16, 2018 · Terminated

NeoBase Succinylacetone Assay Solution, Product Number: 3042-0020; Affected UDI Numbers:…

PerkinElmer Life and Analytical Sciences, Wallac, OY

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: PerkinElmer Life and Analytical Sciences, Wallac, OY · Mustionkatu 6 · Turku, N/A · N/A

Is this the product you have?

  • NeoBase Succinylacetone Assay Solution
  • Product Number3042-0020
  • Affected UDI Numbers(01)06438147295708(17)180531(10)654184 (01)06438147295708(17)180531(10)653414 (01)06438147295708(17)180531(10)653657 (01)06438147295708(17)180630(10)654923 (01)06438147295708(17)180630(10)655680 (01)06438147295708(17)180630(10)655300 (01)06438147295708(17)180630(10)654924 (01)06438147295708(17)180831(10)656025 (01)06438147295708(17)180831(10)657055 (01)06438147295708(17)180831(10)656170 (01)06438147295708(17)180831(10)656320 (01)06438147295708(17)181031(10)657768 (01)06438147295708(17)181031(10)657513 (01)06438147295708(17)181031(10)658006
Lot codes and dates
  • Affected lots
    • 654184
    • 653414
    • 653657
    • 654923
    • 655680
    • 655300
    • 654924
    • 656025
    • 657055
    • 656170
    • 656320
    • 657768
    • 657513
    • 658006
Amount recalled
5275

Where it was sold

Why it was recalled

There is a potential for leaking vials for certain lot numbers where, if proper attire is not used as instructed, a transient minor allergic skin reaction may result when handling leaking succinylacetone reagent vials.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new PerkinElmer Life and Analytical Sciences,… recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 17, 2019 · Data as of Sep 19, 6:15 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.