Reported Jun 10, 2015 · Terminated

Cortical Screw for M/DN Intramedullary Fixation. Metaphyseal/Diaphyseal Nail System, Orthopedic surgical device.

Zimmer, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Zimmer, Inc. · 1800 W Center St · Warsaw, IN · 46580-2304

Is this the product you have?

  • Cortical Screw for M/DN Intramedullary Fixation. Metaphyseal/Diaphyseal Nail System, Orthopedic surgical device.
Lot codes and dates
  • All lots
  • Part numbers00-2253-020-45 INTERLOCKING IM SCREW 4.5MM DIA X 20MM 00-2253-022-45 INTERLOCKING IM SCREW 4.5MM DIA X 22.5MM 00-2253-025-45 INTERLOCKING IM SCREW 4.5MM DIA X 25MM 00-2253-027-45 M/DN Intramedullary Fixation 4.5mm Dia. Cortical Screw 00-2253-030-45 INTERLOCKING IM SCREW 4.5MM DIA X 30MM 00-2253-032-45 INTERLOCKING IM SCREW 4.5MM DIA X 32.5MM 00-2253-035-45 INTERLOCKING IM SCREW 4.5MM DIA X 35MM 00-2253-037-45 INTERLOCKING IM SCREW 4.5MM DIA X 37.5MM 00-2253-040-45 INTERLOCKING IM SCREW 4.5MM DIA X 40MM 00-2253-042-45 INTERLOCKING IM SCREW 4.5MM DIA X 42.5MM 00-2253-045-45 INTERLOCKING IM SCREW 4.5MM DIA X 45MM 00-2253-047-45 INTERLOCKING IM SCREW 4.5MM DIA X 47.5MM 00-2253-050-45 INTERLOCKING IM SCREW 4.5MM DIA X 50MM 00-2253-052-45 INTERLOCKING IM SCREW 4.5MM DIA X 52.5MM 00-2253-055-45 INTERLOCKING IM SCREW 4.5MM DIA X 55MM 00-2253-057-45 INTERLOCKING IM SCREW 4.5MM DIA X 57.5MM 00-2253-060-45 INTERLOCKING IM SCREW 4.5MM DIA X 60MM 00-2253-062-45 INTERLOCKING IM SCREW 4.5MM DIA X 62.5MM 00-2253-065-45 INTERLOCKING IM SCREW 4.5MM DIA X 65MM 00-2253-067-45 INTERLOCKING IM SCREW 4.5MM DIA X 67.5MM 00-2253-070-45 INTERLOCKING IM SCREW 4.5MM DIA X 70MM 00-2253-072-45 INTERLOCKING IM SCREW 4.5MM DIA X 72.5MM 00-2253-075-45 INTERLOCKING IM SCREW 4.5MM DIA X 75MM 00-2253-077-45 INTERLOCKING IM SCREW 4.5MM DIA X 77.5MM 00-2253-080-45 INTERLOCKING IM SCREW 4.5MM DIA X 80MM 00-2253-082-45 INTERLOCKING IM SCREW 4.5MM DIA X 82.5MM 00-2253-085-45 INTERLOCKING IM SCREW 4.5MM DIA X 85MM 00-2253-087-45 INTERLOCKING IM SCREW 4.5MM DIA X 87.5MM 00-2253-090-45 INTERLOCKING IM SCREW 4.5MM DIA X 90MM
Amount recalled
289,616

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Zimmer, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 8, 2015 · Data as of Sep 19, 5:30 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.